Achiziționarea sistemului radiologic digital staționar

Information about customer
Fiscal code/IDNO
Address
5400, MOLDOVA, Rezina, or.Rezina (r-l Rezina), 27 August nr.1
Web site
---
The contact person
Full name
Crijanovschii Eudochia
Contact phone
068469040
Purchase data
Date created
29 Jul 2026, 9:43
Date modified
10 Aug 2026, 8:30
Achizitii.md ID
21661037
CPV
33100000-1 - Echipamente medicale
Type of procedure
Open tender
Award criteria
The lowest price
Funding sources
Advertising
Documents of the procurement procedure
duae.doc
Evaluation Criteria
DUAE
29.07.26 09:43
anunț de participare.pdf anunț de participare.pdf
tenderNotice
Invitație
30.07.26 11:41
ds_bunuri_servicii.docx
Bidding Documents
Documentația Standart
10.08.26 08:30
caiet de sarcini.pdf caiet de sarcini.pdf
Technical Specifications
Caiet de sarcini
10.08.26 08:30
Date:
29 Jul 2026, 10:12
Question's name:
Buna ziua
Question:
va rugam sa incarcati varianta word a caietului de sarcini
Answer (29 Jul 2026, 10:48):
Buna ziua. Caietul de sarcini îl găsiți în Documentația standart modificată.
Date:
29 Jul 2026, 11:12
Question's name:
Termenii și condițiile de livrare/
Question:
Termenii și condițiile de livrare/prestare solicitați: 5 zile de la înregistrare contractului la Trezoreria Regională Centru.va rugam sa modificati in termenii de livrare reali.
Answer (29 Jul 2026, 11:51):
În curînd va fi modificat
Date:
29 Jul 2026, 11:33
Question's name:
DETECTOARE
Question:
Se solicita 2 detectoare *Detector p/u bucky verlical Marimea detector, cm 43x43 Configurare detector : Transmisie prin Cablu da Caracteristicele detectorului Rezolulia matricei, pixeli > 3070 x 3070 ,Adincime imagine : min 16 bit Marimea pixelului <140 um *Detector masa marimea detector (fara fir) Wireles da caracteristicele detectorului : Rezolutia matricei, pixeli min 3070 - 3070 marimea pixelului <140 Pm baterii min2 buc da Nu vedem sens la astfel de configuratie in bucky vertical sa fie detector static(ne mobil cu fir) si in masa sa fie mobil (Wireless) Daca este obligator sa fie 2 atunci: Recomandam sa puneti 2 detectoare statice cu fir : in Vertigraf si Masa Beneficii + Operatorul nu va fi nevoit sa incarce mereu bateriile + Costurile mai mici +Nu este riscul ca detectorul sa fie deteriorat ca de exemplu in cazul unei cazaturi. Ori 2 wireless dar in acest caz va fi unul care se va prafui undeva dar altul se va utiliza la maxim -Costurile vor fi mai mari Reesind din valoarea achizitiei va recomandam sa alegeti ori 1 detector wireless ori 2 cu fir care vor fi plasate in Vertigraf si Masa si vor lucra perfect
Answer (30 Jul 2026, 08:48):
Conform cerințelor solicitate, sunt indicate 2 detectoare digitale. (Varianta cu fir pe tru vertigraf și varianta wireless pentru masa). Este obligatoriu ca detectorul p/u masa sa fie wireless deoarece v-a fi utilizat în diverse proecții oblice, inclusiv direct pe masa sau în diverse poziții ale pacientului. Nu se accepta detector wireless pentru Vertigraf, deoarece, necesita încărcare permanenta a bateriilor și evident schimbul acestora în 4-5 ani. Todotata, conexiunea prin cablu este mai stabila.
Achiziționarea sistemului radiologic digital staționar
Date:
30 Jul 2026, 09:08
Question's name:
Termenii qi condi{iile de livrare/prestare solicitafi: 60 zile de la inregistrare contractului la Trezoreria Regionali Centru
Question:
Buna ziua, luand in considerare termenii reali de producție ai marilor producători pentru echipamente radiologice de acest tip (de regulă, până la 2 luni), solicităm modificarea termenului de livrare de la 60 de zile la 90 de zile de la înregistrarea contractului. Acest termen reprezintă practica uzuală în cadrul licitațiilor naționale și permite participarea unui număr mai mare de producători. Multumim
Answer (30 Jul 2026, 10:51):
Buna ziua. Luând în considerare argumentele tehnice invocate privind termenele reale de producție ale echipamentelor radiologice digitale și în scopul asigurării unei concurențe largi în cadrul procedurii, autoritatea contractantă ACCEPTĂ solicitarea de modificare. Termenul de livrare și instalare a bunului va fi de maximum 90 de zile de la data înregistrării contractului la Trezoreria Regională.”
Date:
4 Aug 2026, 10:07
Question's name:
Afisarea preventiva a imaginii radiologice
Question:
Afisarea preventiva a imaginii radiologice. Solicitam excluderea acestui punct, deoarece este o cerinta restrictiva,  nu are nici o importanta clinica si limiteaza participare altor producatori.
Answer (4 Aug 2026, 14:06):
Bună ziua. Autoritatea contractantă menține această cerință. Afișarea preventivă a imaginii (preview) are o importanță clinică majoră, deoarece permite tehnicianului să verifice instantaneu poziționarea corectă a pacientului și să evite iradierea repetată inutilă a acestuia. Funcția reprezintă un standard tehnologic actual, oferit de majoritatea producătorilor de echipamente radiologice digitale, prin urmare cerința NU este restrictivă. Specificațiile tehnice rămân neschimbate.
Date:
4 Aug 2026, 10:07
Question's name:
Lungimea suprafeței mesei ≥230
Question:
Lungimea suprafeței mesei ≥230: Este caracteristica doar anumitor producator. Solicitam modificarea acestui parametru in: Lungimea suprafeței mesei ≥230+/- 5 %.
Answer (5 Aug 2026, 09:59):
Buna ziua. Se acceptă propunerea. Parametrul tehnic se va modifica.
Date:
4 Aug 2026, 10:08
Question's name:
Frecventa mximala, kHz min 200
Question:
Frecventa mximala, kHz min 200: deoarece este un parametru restrictiv, si nu are nici o importanta la calitatea examinarii, solicitam excluderea sau modificarea in Generator de Frecventa Inalta.
Answer (5 Aug 2026, 10:04):
Buna ziua. Se acceptă modicicarea în scopul extinderii concureenței. Parametrul tehnic se va modifica.
Date:
4 Aug 2026, 10:08
Question's name:
Miscare verticala, cm pozitia de jos  max 50cm
Question:
Miscare verticala, cm pozitia de jos  max 50cm: este un parametru restrictiv, solicitam modificarea in: Miscare verticala, cm pozitia de jos  max 50cm +/- 5 %
Answer (5 Aug 2026, 10:05):
Buna ziua. Se acceptă propunerea. Parametrul tehnic se va modifica.
Date:
4 Aug 2026, 10:09
Question's name:
Deplasarea longitudinală sistem Bucky al mesei, cm 50
Question:
Deplasarea longitudinală sistem Bucky al mesei, cm 50: este un parametru restrictiv, solicitam modificarea in: Deplasarea longitudinală 50cm +/- 5 %
Answer (5 Aug 2026, 10:06):
Buna ziua. Se acceptă propunerea. Parametrul tehnic se va modifica.
Date:
4 Aug 2026, 10:09
Question's name:
Deplasarea verticala bucky, cm ≥ 150 cm
Question:
Deplasarea verticala bucky, cm ≥ 150 cm: restrictioneaza participarea unor producatori. Solicitam modificarea in Deplasarea verticala bucky, cm ≥ 150 cm +/- 5 cm
Answer (5 Aug 2026, 10:07):
Buna ziua. Se acceptă propunerea. Parametrul tehnic se va modifica.
Date:
4 Aug 2026, 10:09
Question's name:
Rotire suspensie : ± 900
Question:
Rotire suspensie : ± 900: nu este clar la ce se refera acest punct. Probabil s-a avut in vedere Grade, iar atunci ar trebui modificat in: Rotire suspensie : ± 90 grade
Answer (5 Aug 2026, 10:10):
Buna ziua. Se confirmă o eroare de dactilografiere în textul inițial. Parametrul corect este: Rotire suspensie: ± 90 grade (± 90°)”.
Date:
4 Aug 2026, 14:11
Question's name:
sistem de radiografie digitală cu stativ suspendat de tavan (ceiling-mounted) sau un sistem cu stativ fixat pe podea (floor-mounted).
Question:
Vă rugăm să clarificați și să specificați în caietul de sarcini tipul sistemului radiologic solicitat, respectiv dacă se dorește un sistem de radiografie digitală cu stativ suspendat de tavan (ceiling-mounted) sau un sistem cu stativ fixat pe podea (floor-mounted). În forma actuală, documentația de atribuire nu precizează această caracteristică, ceea ce poate conduce la interpretări diferite din partea operatorilor economici și la prezentarea unor oferte care nu sunt comparabile din punct de vedere tehnic și financiar. Totodată, menționăm că sistemele cu stativ suspendat de tavan sunt, în mod obișnuit, considerabil mai costisitoare decât sistemele cu stativ fixat pe podea. Având în vedere bugetul estimativ al procedurii, apreciem că achiziționarea unui sistem suspendat de tavan ar putea depăși valoarea disponibilă. Prin urmare, vă rugăm să precizați în mod expres în caietul de sarcini tipul de sistem radiologic solicitat, pentru a asigura elaborarea unor oferte conforme și comparabile. Multumim
Answer (5 Aug 2026, 10:17):
Bună ziua. Autoritatea contractantă clarifică faptul că se solicită un sistem radiografic digital cu stativul tubului radiologic cu fixare pe podea (floor-mounted).Această opțiune este necesară pentru a asigura compatibilitatea cu infrastructura actuală a instituției medicale și pentru a evita lucrări suplimentare de consolidare structurală a tavanului cabinetului. Echipamentul ofertat trebuie să asigure toate deplasările și parametrii solicitați (deplasare longitudinală min. 250 cm, deplasare verticală 40-189 cm și urmărire automată/tracking). Specificațiile tehnice rămân neschimbate.
Date:
4 Aug 2026, 16:09
Question's name:
Afisarea preventiva a imaginii radiologice
Question:
Răspuns (4 aug 2026, 14:06): din raspunsul oferit nu este clar la ce se refera aceasta optiune: Previzualizarea (Preview) imaginii radiografice sau la Afișarea imaginii pentru verificarea poziționării pacientului. Va rugam sa specificati
Answer (7 Aug 2026, 14:42):
Parametrul solicitat se refera la viteza de afișare a imaginii radiologice pe stația de achiziție. Valoarea maxima solicitata pentru afisare imagine de max 5 sec. reprezinta timpul necesar pentru transfer, procesare si afisarea imaginii provenite de la detectorul digital, fiind un indicator al performantei intregului lant de achizitie a imaginii. Acest parametru este stabilit prin arhitectura hardware si software a sistemului. Astfel parametrul nu se modifica.
Date:
4 Aug 2026, 16:18
Question's name:
Afișarea imaginii
Question:
Afișarea imaginii pe ecranul de la tub este o funcție de confort (workflow), nu una esențială pentru performanța sistemului. Insistam ca aceasta functie este restrictivă, deoarece elimină sisteme care oferă aceeași performanță clinică, dar afișează imaginea pe consola de achiziție sau pe o consolă mobilă, nu pe ecranul tubului. Mai ales ca expunerile se fac de la consola de achizitie iar tehnicianul poate verifica instantaneu imaginea, fara a fi obligat sa se apropie de tubul radiologic. În acest caz, este rezonabil să solicitam modificarea cerinței într-o formulare mai neutră, de exemplu: „Sistemul trebuie să permită previzualizarea imaginii radiografice imediat după achiziție pe consola de achiziție și/sau pe ecranul tactil al tubului radiologic.”
Answer (7 Aug 2026, 14:43):
Afisarea imaginii pe ecranul tactil al tubului radiologic reprezinta o cerinta esentiala, nu doar o functie de confort. Aceasta functionalitate contribuie direct la eficienta si siguranta fluxului de lucru, permitind tehnicianului radiolog sa verifice imediat calitatea imaginii si pozitionarea pacientului fara a fi necesara deplasarea la consola de achizitie. In acelasi timp, pacientul poate fi sigur de investigatia efectuata si respectiv doza de iradiere obtinuta. In situatiile clinice in care sunt necesare ajustari rapide s-au repetarea expunerii , accesul instantaneu la imagine reduce timpul de examinare și contribuie la limitarea expunerii inutile. Mai mult, in cazul sistemelor radiologice moderne, afisarea imaginii pe ecranul tactil al tubului este o functionalitate standard. Din aceste motive ca aceasta cerinta trebuie mentinuta ca obligatorie.
Date:
6 Aug 2026, 14:09
Question's name:
Miscare verticala, cm pozitia de jos max 50cm- pozitia de sus max 85cm
Question:
Solicitam sa modificati astfel Miscare verticala, cm pozitia de jos minim 50cm- pozitia de sus minim 85cm. Multumim
Answer (10 Aug 2026, 08:10):
Nu se accepta solicitarea, deoarece ne dorim ca miscarea pe verticala a mesei in pozitia de jos sa nu depaseasca 50 cm, si respectiv 85 cm. Specificarea valorilor maxime stabileste limita dimensionala a echipamentuluisi garanteaza accesibilitatea pentru toate categoriile de pacienti si utilizatori, inclusive in cazul examinarii pacientilor cu mobilitate redusa s-au aflati pe targa. Inlocuirea cerintei cu minim 50cm si minim 85cm, v-a permite ofertarea unor sisteme care nu pot cobori sufficient s-au care depasesc in mod nejustificat inaltimea necesara, afectind ergonomia, siguranta pacientului si flexibilitatea clinica.
Date:
6 Aug 2026, 14:10
Question's name:
Densitatea mesei <0,6 mm Al (Fibra de carbon)
Question:
Solicitam sa modificati astfel Densitatea mesei ≤0,7 mm Al (Fibra de carbon). Multumim
Answer (10 Aug 2026, 08:11):
Cerinta privind densitatea (Echivalentul de atenuare) a mesei <0,6 mm Al, nu se modifica deoarece aceasta reprezinta un parametru essential care influenteaza direct calitatea imaginii radiologice si doza de iradiere administrate pacientului. Modificarea cerintei la ≤0,7 mm Al, ar permite ofertarea unor echipamente cu o atenuare mai mare a fascicului de raze X, ceeea ce necesita majorarea parametrilor de expunere pentru a obtine aceeasi calitate a imaginii.
Date:
6 Aug 2026, 14:14
Question's name:
Miscarea mesei Longitudinala, ≥90cm Laterala, ≥25cm
Question:
Rugam sa specificati la care parametri exacti faceti referitna, deoarece miscare mesei in plan longitudinal pentru un x-ray fixat pe podea deobicei are valori ≥43cm, iar in plan transversal ≥10cm. Multumim
Answer (10 Aug 2026, 08:12):
Cerinta se refera la cursa de deplasare a blatului mesei, respectiv deplasare longitudinala de la izocentru in stinga si in dreapta. Cursa totala de deplasare longitudianala stinga -dreapta se solicita ≥90cm. Respectiv si cursa transversala de la izocentru dispune de miscare stinga-dreapta in total de ≥25cm. Valorile mentionate de Dvs, reprezinta cursa intr-o singura directie. Respectiv cerinta nu v-a fi modificata.
Date:
6 Aug 2026, 14:17
Question's name:
Dimensiunea, cm >43x43
Question:
Solicitam sa modificati astfel Dimensiunea, cm >42x42. Multumim
Answer (10 Aug 2026, 08:13):
Dimensiunea solicitata >43x43, este o una standard, ce face referinta la dimensiunea Bucky din masa radiologica si bucky vertical. Acesta este spatial destinat pentru detectorul digital. O dimensiune mai mica nu v-a pemite inadrarea detectoarelor solicitate. Astfel parametru nu se modifica
Date:
6 Aug 2026, 14:20
Question's name:
Controlul automat al expunerii (AEC) min 3 cimpuri
Question:
Rugam sa modificati astel Controlul automat al expunerii (AEC) - obligatoriu
Answer (10 Aug 2026, 08:13):
Controlul automat al expunerii (AEC) cu minim 3 cimpuri de masurare nu se modifica deoarece numarul de camere AEC reprezinta un parametru tehnic essential care influenteaza flexibilitatea si precizia controlului expunerii. Configuratia cu minim 3 cimpuri AEC reprezinta standardul utilizat in sistemele radiologice moderne, care permitea selectarea optima a zonei de masurare in functie de regiunea anatomica examinata.
Date:
6 Aug 2026, 14:25
Question's name:
Deplasarea longitudinald, cm 50
Question:
Clarificati va rog daca se face referire la masa sau la Bucky wall stand
Answer (10 Aug 2026, 08:14):
Multumim pentru clarificare. Deplasare longitudinala de 50 cm face referinta la bucky din masa radiologica
Date:
6 Aug 2026, 14:36
Question's name:
Confrgurare detector : Transmisie prin Cablu da
Question:
Rugam sa modificati astfel Confrgurare detector : Transmisie prin Cablu sau Wireless
Answer (10 Aug 2026, 08:15):
Conform cerințelor solicitate, sunt indicate 2 detectoare digitale. (Varianta cu fir pe tru vertigraf și varianta wireless pentru masa). Este obligatoriu ca detectorul p/u masa sa fie wireless deoarece v-a fi utilizat în diverse proecții oblice, inclusiv direct pe masa sau în diverse poziții ale pacientului. Nu se accepta detector wireless pentru Vertigraf, deoarece, necesita încărcare permanenta a bateriilor și evident schimbul acestora în 4-5 ani. Todotata, conexiunea prin cablu este mai stabila
Date:
6 Aug 2026, 14:38
Question's name:
baterii min2 buc da
Question:
Conform producatorilor detectorul se incarca nu vine configurat cu baterii de schiimb, rugam sa excludeti acest punct
Answer (10 Aug 2026, 08:16):
In configuratia solicitata, este indicat expres ca detectorul wireless pentru masa radiologica sa dispuna de baterii de schimb-minim 2 buc. Probabil este cunoscut faptul ca durata de viata a unei baterii este limitata la citeva mii de cicluri de incarcare. In cazul Dvs, v-a fi necesar de schimbat detectorul digital in totalitate, ceea ce este foarte costisitor. Este o cerinta obligatorie si nu poate fi modificata.
Date:
6 Aug 2026, 14:39
Question's name:
autonomie min 8 ore
Question:
Solicitam sa modificati autonomie min 7 ore
Answer (10 Aug 2026, 08:16):
Autonomia de 8 ore este obligatoriu, deoarece aceasta v-a asigura continuitatea activitatii clinic pe durata unui schimb complet de lucru, fara intreruperi cauzate de necesitatea incarcarii s-au inlocuirii bateriei. Reducerea la 7 ore diminueaza rezerva operationala si poate afecta continuitatea activitatii echipamentului radiologic. Parametru este obligatoriu si nu se modifica.
Date:
6 Aug 2026, 14:50
Question's name:
Capacitatea termicd, KHU >300
Question:
Solicitam sa modificati astfel Capacitatea termica, KHU >200
Answer (10 Aug 2026, 08:17):
O capacitate termica de 300 KHU, permite desfasurarea in conditii optime a unui volum ridicat de investigatii, reducind timpul de asteptare determinati de racirea anodului. Reducerea cerintei diminueaza rezerva termica a tubului, ceea ce poate duce la intreruperi frecvente si limitarea perormantei in exploatarea clinica. Parametrul nu se modifica
Date:
6 Aug 2026, 15:01
Question's name:
Deplasarea longitudinală a suspensiei tubului ≥250 cm
Question:
Solicitam sa modificati astfel Deplasarea longitudinală a suspensiei tubului ≥150 cm. Conform specifica'iiloe tehnice a marilor producatori pentru x-ray fixat pe podea, nu pot fi atinse valorile solicitat epentru deplasarea longitudinala. Multumim
Answer (10 Aug 2026, 08:18):
Deplasarea longitudianala a suspensiei tubului de minim 250 cm nu se modifica deoarece este necesar pentru a fi asigurata acoperirea completa a mesei radiologice si respectiv bucky vertical, precum si efectuarea tuturor tipurilor de examinari fara limitari de pozitionare. Reducerea la 150 cm ar limita semnificativ domeniul de deplasare a suspensiei tubului si respectiv v-a fi necesar de repozitionat pacientul.
Date:
6 Aug 2026, 15:03
Question's name:
depiasarea verticala suspensie tub radiologic (scanare orizontala), cm min 40-189 cm
Question:
rugam sa analizati posibilitatea de a modifcia astfel depiasarea verticala suspensie tub radiologic (scanare orizontala), cm min 40-130 cm, pentru x-ray fixat pe podea nu pot fi atinse valorile stabilite, informatii sustinunte de cate toti producatorii prezenti.
Answer (10 Aug 2026, 08:18):
Deplasarea verticala a suspensiei tubului radiologic in intervalul minim 40-189 cm nu se modifica, deoarece aceasta asigura posibilitatea efectuarii intregii game de examinari radiologice, inclusive la pacientii in pozitie ortostatica, pe targa s-au scaun cu rotile, fara limitari de pozitionare. Intervalul solicitat ofera flexibilitatea necesara alinierii tubului radiologic cu masa si bucky-ului vertical.
Date:
10 Aug 2026, 19:41
Question's name:
Monitor luminos cu min. 1000 CD/m2 (mii jale de bani bugetari)
Question:
*Monitor medical > 30" inci -e posibil de gasit! *(Rezolutie min 3800 x 2100) mai greu dar ok ( numai nu cu cablu VGA va rog) ****Luminozitatea min. 1000 CD/m2 Valoare restrictiva!!! ***** Va aducem la cunostinta diagnostic radiologic avansat utilizeaza de la 350–700 cd/m² pe cind 1000CD/m2 sunt monitoare HIGH-END si se folosesc in Mamografie si la CT,RMN Pretul la astfel de monitoare este > 5000Euro Reiesind din valoarea estimata va rugam sa micsorati valoare pina la ">= 500 CD/m2 ori 500 nits" pentru a nu limita participarea la aceasta licitatie.
Answer (12 Aug 2026, 08:17):
Cerinta de minim 1000 CD/m2 a fost stabilita pentru a asigura un nivel ridicat de performanta, destinat pentru afisarea imaginilor radiologice in scop diagnostic. Nivelul ridicat de luminozitate ofera variatii fine de densitate, absolut necesar pentru investigatiile radio-diagnostice. Cerinta nu urmareste limitarea concurentei, dar stabilirea unui nivel minim de performanta , corespunzator utilizarii monitorului in cadrul unui system de imagistica medicala. Parametrul nu se modifica
Date:
11 Aug 2026, 10:42
Question's name:
Afișarea imaginii
Question:
Înțelegem argumentele prezentate cu privire la utilitatea afișării imaginii pe ecranul tactil al tubului radiologic. Cu toate acestea, considerăm că această funcționalitate nu trebuie stabilită ca cerință obligatorie în cadrul specificației tehnice. Afișarea imaginii pe ecranul tactil al tubului reprezintă o funcționalitate suplimentară, care poate contribui la confortul și eficiența fluxului de lucru, însă nu constituie o cerință tehnică obligatorie pentru funcționarea în condiții de siguranță și pentru realizarea corespunzătoare a investigațiilor radiologice. Verificarea imaginii, a calității acesteia și a poziționării pacientului se poate efectua la consola de achiziție, care este destinată în mod specific acestor operațiuni. De asemenea, apreciem că includerea acestei funcționalități ca cerință obligatorie poate limita în mod nejustificat gama de echipamente care pot răspunde cerințelor tehnice și clinice ale sistemului, fără ca acest lucru să reprezinte în mod necesar un avantaj esențial în procesul de diagnostic. În consecință, considerăm oportună eliminarea acestei cerințe din categoria cerințelor obligatorii. Funcționalitatea poate fi menționată, dacă se consideră necesar, ca facilitate suplimentară/opțională, fără condiționarea conformității echipamentului de existența acesteia.
Answer (12 Aug 2026, 08:18):
În cazul unui sistem radiologic digital modern, afișarea imediată a imaginii la nivelul tubului radiologic are un rol important în optimizarea fluxului de lucru, verificarea imediată a rezultatului expunerii și reducerea necesității deplasării operatorului între poziția de examinare și consola de achiziție. Această funcționalitate este deosebit de relevantă în situațiile în care operatorul efectuează examinarea direct lângă pacient și trebuie să verifice imediat dacă poziționarea, colimarea și acoperirea anatomică sunt corespunzătoare. Accesul imediat la imagine permite identificarea rapidă a unei eventuale necesități de repetare a examinării și contribuie la evitarea unor deplasări și operațiuni suplimentare în cadrul fluxului de lucru. Menționăm că solicitarea nu presupune că imaginea de pe ecranul tubului înlocuiește consola de achiziție sau stația de diagnostic. Consola rămâne destinată procesării și evaluării complete a imaginii, în timp ce afișarea imediată pe ecranul tubului reprezintă o componentă funcțională a fluxului de achiziție, care permite operatorului să verifice rapid rezultatul expunerii. În plus, cerința este formulată în scopul obținerii unui nivel minim de funcționalitate și eficiență operațională a sistemului, nu pentru favorizarea unui anumit producător. Prin urmare, considerăm că eliminarea cerinței și transformarea acesteia într-o facilitate opțională ar conduce la diminuarea nivelului funcțional minim solicitat pentru echipament și ar permite ofertarea unor sisteme cu un nivel inferior de integrare și ergonomie. Astfel menținem cerința privind « afișarea imaginii radiografice pe ecranul tactil al tubului radiologic», ca o componenta obligatorie a sistemului.
Date:
11 Aug 2026, 17:49
Question's name:
Geam plumbat min 1000x800 mm cu rama radioprotectoare
Question:
Vă rugăm să reanalizați cerința privind includerea în ofertă a geamului plumbat cu dimensiunea minimă de 1000 × 800 mm, cu ramă radioprotectoare. Menționăm că un astfel de geam plumbat nu a fost solicitat în mod uzual în cadrul licitațiilor publice pentru echipamente de radiologie. În acest context, pentru a evita includerea unor componente care nu sunt necesare în configurația standard a echipamentului și pentru a permite o participare cât mai largă a operatorilor economici, vă rugăm să excludeți această cerință din configurația obligatorie.
Answer (12 Aug 2026, 08:16):
Geamul plumbat reprezintă o componentă necesară a sistemului de radioprotecție, asigurând protecția personalului și, concomitent, posibilitatea supravegherii vizuale a pacientului și echipamentului din camera de comandă, fără întreruperea barierei de ecranare. Prin urmare, cerința privind geamul plumbat rămâne neschimbată.