1
Enquiry period
with 02.05.2025 13:54
to 09.06.2025 14:38
25 days left
2
Bidding period
with 09.06.2025 14:38
to 30.06.2025 14:38
3
Auction
will not be used
4
Evaluation

5
Contract

Status Enquiry period
Estimated value without VAT 9 424 465,34 MDL
Period of clarifications: 2 May 2025, 13:54 - 9 Jun 2025, 14:38
Submission of proposals: 9 Jun 2025, 14:38 - 30 Jun 2025, 14:38

Supplier technical support:

(+373) 79999801


This procedure is carried out without auction. Your offer is final and must contain the entire list of required documents.

Advertising
Subscribe
Advertising

Conform Anunțului de participare

Information about customer
Fiscal code/IDNO
Address
MD2004, MOLDOVA, mun.Chişinău, mun.Chişinău, bd. Ștefan cel Mare 165
Web site
---
The contact person
Full name
Eugenia Gligor
Contact phone
022205265
Purchase data
Date created
2 May 2025, 13:54
Date modified
13 May 2025, 11:18
Achizitii.md ID
21413944
CPV
33100000-1 - Echipamente medicale
Type of procedure
Open tender
Award criteria
The lowest price
Funding sources
List of lots
Lot nr. 22 - 22. Spirometru
Budget: 9166.67 MDL
Active
Lot nr. 23 - 23. Tonometru
Budget: 3750.0 MDL
Active
Lot nr. 24 - 24. Fonendoscop
Budget: 1833.33 MDL
Active
Lot nr. 27 - 27. Pulsoximetru
Budget: 6666.67 MDL
Active
Lot nr. 30 - 30. Negatoscop
Budget: 4625.0 MDL
Active
Lot nr. 44 - 44. Luxmetru
Budget: 46265.0 MDL
Active
Lot nr. 47 - 47. Urodensimetru
Budget: 17000.0 MDL
Active
Lot nr. 55 - 55. Dulap medical
Budget: 40500.0 MDL
Active
Lot nr. 58 - 58. Portac
Budget: 4450.0 MDL
Active
Lot nr. 60 - 60. Foarfece
Budget: 5875.0 MDL
Active
Lot nr. 78 - 78. Emplastru
Budget: 2500.0 MDL
Active
Lot nr. 80 - 80. Electrozi
Budget: 2500.0 MDL
Active
Lot nr. 94 - 94. Hotă chimică
Budget: 80333.33 MDL
Active
Advertising
Documents of the procurement procedure
declarație beneficiar.doc
Bidding Documents
-
2.05.25 13:54
duae.doc
Bidding Documents
-
2.05.25 13:54
documentația standard 115.docx
Bidding Documents
-
12.05.25 14:38
anexa nr. 2 anunț de participare.docx
Bidding Documents
-
12.05.25 14:38
32. Lampa bactericidă medicală recirculară de perete
Date:
2 May 2025, 19:56
Question's name:
Lot 32
Question:
Ce înseamnă "Nivelul potențialului risc 2 E ?
Answer (5 May 2025, 08:25):
Nivelul de risc 2 E reprezinta distrugera a microorganisme care pot cauza boli (ex: virusul gripal comun).
32. Lampa bactericidă medicală recirculară de perete
Date:
2 May 2025, 20:03
Question's name:
Lot 32
Question:
Buna seara, Dumneavoastră indicati clasa de risc 1, dar indicați nivel de protecție de la percutare tip B, Întrebarea constă : aveti prevăzut pentru conectarea fiecărui Recirculator Automat separat. În alt caz clasa 1 prevede cablu utilajului 3 fire, tip B e utilaj cu 2 fire si necesită asigurarea de către instituție a măsurilor de protecție suplimentară. Este gata ingineri sași asume riscul în caz de percutare electrică a angajaților ?????
Answer (5 May 2025, 08:25):
Tip B este utilizat pentru a preveni electrocutarea și incendiile cauzate de curenți de scurgere. Cablu cu 3 fire, impămintare obligatorie, și protectie de la supratensiune sau scurt circuit, Producătorul asigură funcționarea corectă a dispozitivului medical emițând declarația de conformitate care este aprobată prin înregistrarea în Registrul de Stat al Dispozitivelor Medicale.
32. Lampa bactericidă medicală recirculară de perete
Date:
2 May 2025, 20:06
Question's name:
Lampa Recirculara
Question:
În condiții nu se întâlnește cerința legată de efectu bactericid , sau nu-i necesar ? Cumpărați sa fie cumpărat și ceva sa bizie pe perete.
Answer (5 May 2025, 08:28):
Buna ziua. Descrierea corespunde necesităților actuale ale autorității contractuale
75. Analizator semiautomat de biochimie
Date:
5 May 2025, 09:13
Question's name:
Software
Question:
Bună ziua! Software-ul pe care îl solicitați este utilizat pentru mașinile unui anumit producător - Spectrolab. Această cerință reduce concurența, vă rugăm să scoateți acest punct și să țineți cont de faptul că agenții economici vor oferi analizatoare cu propriul software, care este proiectat pentru acestea. Vă mulțumim anticipat!
Answer (5 May 2025, 11:28):
Buna ziua. Descrierea este făcută conform necesităților Autorității contractante. Dacă produsul Dvs este analogic cu cerințele noastre și primul la preț, veți fi selectat ca cîștigător.
Date:
5 May 2025, 09:20
Question's name:
Documentație de atribuire
Question:
Bună ziua! Vă rugăm să clarificați, dacă certificația Ce este obligatorie pentru fiecare lot? Vă mulțumim!
Answer (5 May 2025, 11:29):
Buna ziua. În specificații tehnice la fiecare lot este indicat ce fel de certificate se solicită. Dacă nu se solicită, înseamnă că nu e obligatoriu.
21. Aparat ultrasonograf
Date:
5 May 2025, 11:04
Question's name:
Specificatie
Question:
Buna ziua, in specificatie este mentionat producatorul si modelul exact. Va rugam sa excludeti patrametrii care fac referire la un anumit producator.
Answer (5 May 2025, 11:38):
Buna ziua. Mulțumim pentru colaborare.
75. Analizator semiautomat de biochimie
Date:
5 May 2025, 11:58
Question's name:
Descrierea
Question:
Descrierea dvs. este haotică și cu două sensuri. Aveți fiecare cerință tehnică dublată cu o descriere diferită (ex. Ecran LCD mare, ușor de citit, și se dublează Display: LCD 128x64). Această descriere este înșelătoare și poate cauza neînțelegeri din partea operatorilor economici. Vă rugăm să revizuiți, să editați și să faceți descrierea acestui lot mai clară pentru operatorii economici.
Answer (5 May 2025, 13:57):
Buna ziua. Display: LCD nu mai mic de 128x64
75. Analizator semiautomat de biochimie
Date:
5 May 2025, 12:07
Question's name:
Clarificare
Question:
Utilizarea unui parametru în cerințele tehnice care indică un anumit producător (în acest caz, denumirea comercială a software-ului unui anumit producător de analizoare de tipul solicitat), care subminează principiile licitațiilor deschise, reducând concurența. Dacă solicitați un analizor, sunteți obligat să specificați cerințe tehnice de tip generalizat. Aveți în descriere "Aplicație software care oferă un control complet al funcțiilor spectrofotometrului de la un computer și permite efectuarea scanării, analiza multilungimi de unda si ADN/proteine." - este o descrierea acceptabilă, dar "Program software MWave basic" - este una incorectă și neîntemeiată. Solicităm scoaterea ei. Și cum a fost menționat în clarificare de mai sus - descrierea Dvs. necesită modificări, căci induce în eroare prin dublarea punctelor tehnici.
Answer (5 May 2025, 14:01):
se substituie cu Program software care susține integral toate funcționalitățile echipamentului în condiții de activitate maximă
75. Analizator semiautomat de biochimie
Date:
5 May 2025, 14:27
Question's name:
Descrierea
Question:
Display, Marimi analizorului - De asemenea, acest parametru nu reflectă formarea corectă a unei descrieri generalizate a aparatului. Specificați o dimensiune specifică, care este dimensiunea aparatelor unei familii de aparatele producătorului SPECTROLAB, încălcând din nou principiul licitației, favorizând astfel un anumit model și un anumit producător. Descrierea dvs. este pur și simplu o descriere a unui dispozitiv SPECTROLAB specific, vă rugăm să consultați următoarele linkuri - https://spectrolab.com.ua/ua/p19134156-spektrofotometr-1100d.html , https://spectrolab.com.ua/ua/p19143422-spektrofotometr-1100.html, https://spectrolab.com.ua/ua/p19164274-spektrofotometr-1200.html, https://spectrolab.com.ua/ua/p19162100-spektrofotometr-1200.html . Aproape toată descrierea este preluată de aici. Acest lucru arată în mod clar o înțelegere între dvs. și un agent economic, care încalcă legea. Solicităm insistent anularea acestui lot și editarea completă a descrierii tehnice cu publicare unei descrieri generice, care să nu încalce principiile licitațiilor și legislația privind licitațiile!
Answer (5 May 2025, 15:36):
Buna ziua. În descriere nu este indicat producătorul și dimensiunile survin din necesitățile laboratorului. Dacă veniți cu oferta prima la preț cu devieri neesențiale, dar mai bune ale specificațiilor tehnice, oferta Dvs poate fi cîștigătoare.
75. Analizator semiautomat de biochimie
Date:
6 May 2025, 08:22
Question's name:
75. Analizator semiautomat de biochimie
Question:
Daca doriti concurenta în licitatia la acest lot, dar nu mimare de procedura deschisa de achizitie, va rugam sa eliminati/corectati puncturi cu ecranul și dimensiuni, deoarece cerințele indicate de Dvs. a utilajului nu vorbesc despre performanța tehnica a utilajului medical, favorizind doar unului producător, care poate încadra în cerințele neîntemeiate. Vă rugăm să răspundeți și să acționați concret, dar nu să utilizați frazele tipice, aveți o descriere incorectă a mașinii solicitate, care încalcă principiile legii - articolul 7, alineatele c, h, j, k. Sperăm în respectul dumneavoastră reciproc față de agenții economici. Mulțumim anticipat.
Answer (6 May 2025, 09:48):
Buna ziua. Descrierile tehnice sunt conform necesităților Auitorității contractante. Ceea ce se referă la ecran, trebuie să aibă dimensiunile minime128x64. Nu sunteți limitat în dimensiuni mai mari.
Date:
6 May 2025, 10:24
Question's name:
Lot 20 Aparat p/u testarea cardiopulmonara
Question:
Specificatia data este sub producatorul MES, model VO2max Finder. Se solicita modificarea acesteia pentru ca si alti agenti economici sa poata participa. Se solicita eliminarea parametrilor: Masurarea debitului-''Rata flux:2.4 ml/sec'', ''Ventilatia masurata: 300 l / min'' , ''Precizie: < 2 ''. Se solicita modificarea parametrilor: Analizator de oxygen, ''Domeniul de masurare: 0 - 25%'' in ''Domeniul de masurare: 10 - 25%'' (fiindca pana la 10% masurile sunt irelevante), ''Timp de raspuns: t 90 < 50 ms'' in ''Timp de raspuns: t 90 < 200 ms'', ''Precizie: ± 0,02%'' in ''Precizie: ± 0,02%'' in ''Precizie: ± 0,1%''; ''Analizator de dioxid de carbon: absorbtie infrarosu NDIR'' in ''Analizator de dioxid de carbon: absorbtie infrarosu NDIR sau tehnologie alternativa'', ''Timp de raspuns: t90 <90 ms'' in ''Timp de raspuns: t90 <100 ms'', ''Rezolutie: 0,01%'' in ''Rezolutie: 0,1%''; ''Interfata de conectare: USB si RS232'' in ''Interfata de conectare: USB sau RS232'' Rog sa se modifice specificatia tehnica nr.13: ''Masurarea debitului: Interval de debit: ± 20 l / s Rata flux: 2,4 ml / sec Debit minim : 10 ml / s'' in ''Masurarea debitului: Interval de debit: ± 18 l / s Rata flux: 50 ml / sec Debit minim : 10 ml / s'' pentru a permite participarea mai multor agenti economici. Specificatii din anexa limiteaza acest fapt. Va multumesc Se solicita modificarea Domeniului de masurare a volumului (0-20)L in (0-18)L pentru a permite participarea mai multor producatori fara a limita capacitatile dispozitivului. Multumesc anticipat
Answer (6 May 2025, 10:43):
Buna ziua. Specificațiile tehnice se scriu conform necesităților cumpărătorului, dar nu obiectelor ce le au la depozit unii vînzători. Vă amintim că conform Articolului 37. Regulile privind descrierea bunurilor, lucrărilor şi serviciilor (1) Specificațiile tehnice ale bunurilor, lucrărilor şi serviciilor solicitate de autoritatea contractantă vor reprezenta o descriere exactă şi completă a obiectului achiziţiei, astfel încît fiecare cerinţă şi criteriu, stabilite de autoritatea contractantă, să fie îndeplinite.
Date:
6 May 2025, 14:49
Question's name:
Lot 20 Aparat p/u testarea cardiopulmonara
Question:
Buna ziua. Specificatiile tehnice nu sunt scrise conform necesitatilor cumparatorului, dar conform specificatiilor unui singur produs. Specificatii tehnice sunt exact ale producatorului MES, model VO2max Finder. Autoritatea beneficiara este obligata ca specificatii tehnice sa fie intocmite astfel incat mai multe produse sa se incadreze. Specificatiile tehnice nu reprezinta un proces de copy paste al specificatiilor unui singur produs. Nemodificarea specificatiilor vor duce la depunerea unei contestatii. Multumesc!
Answer (6 May 2025, 15:11):
Buna ziua. Autoritatea contractantă nu are specialiști tehnici, ci medici care au nevoie de utilaje conform necesităților prezente. Aparatul dat are cerințe ce trebuie să aibă. Dacă găsiți mențiunea directă a unui producător, vă rog ne indicați și eliminăm eroarea.
Date:
6 May 2025, 16:14
Question's name:
Lot 20 Aparat p/u testarea cardiopulmonara
Question:
Drept dovada ca specificatiile tehnice pentru acest lot reprezinta o copie a specificatiilor din data sheet, se ataseaza urmatorul link, unde acesta poate fi gasit. https://mes.com.pl/en/produkty/ergospirometers/ergospirometer-vo2max-finder/ Specificatiile incepnad cu punctul 13 pana la 20 reprezinta o copie exacta. Nici un alt dispozitiv nu va avea exact aceeasi parametri. Se atrage atentia la punctul 20: Patente utilizate: Patent 173767, Patent 195041, Patent 230143. Niciun alt producator nu le prezinta in datasheet, mai ales in aceeasi ordine. Se solicita respectuos, analiza si modificarea parametrilor incepand cu punctul 13 pana la 20. Ca reper, solicit sa analizati sistemele de la Cosmed, Schiller, BTL, dar si alti producatori. Multumesc
Answer (6 May 2025, 16:58):
Buna ziua. Dacă găsiți mențiunea directă a unui producător, vă rog ne indicați și eliminăm eroarea.
Date:
7 May 2025, 12:57
Question's name:
Anunt de participare
Question:
Anexati anuntul de participare. Multumim
Answer (7 May 2025, 13:29):
Buna ziua Stimate operator economic, Anuntul de participare (Anexa 2) este anexat pe platforma electronica O zi buna
3. Cuptor de calcinare
Date:
7 May 2025, 13:41
Question's name:
Cuptor de calcinare
Question:
Buna ziua. Un cuptor de calcinare NABERTHERM are pretul de minim 3,5 ori mai mare decat bugetul estimat lotului. Va rugam frumos sa revizuiti bugetul pentru acest lot. Va multumim.
Answer (7 May 2025, 13:56):
Buna ziua. Acesta e bugetul estimat de autoritatea contractanta pentru lotul respectiv O zi buna
93. Agitator-rotator programabil cu suport pentru tuburi de volum 1.5 – 16 ml
Date:
7 May 2025, 14:33
Question's name:
93. Agitator-rotator programabil cu suport pentru tuburi de volum 1.5 – 16 ml ±1 ml
Question:
Stimate autoritateă adresăm respectuos solicitarea de a analiza posibilitatea modificării specificațiilor tehnice aferente procedurii de achiziție a lotului 93. Agitator-rotator programabil cu suport pentru tuburi de volum 1.5 – 16 ml, în sensul introducerii unei marje de toleranță pentru limita superioara de 16 ml ±1 ml pentru volumul nominal al tuburilor. Considerăm că această modificare nu afectează scopul achiziției și este în conformitate cu principiile concurenței și eficienței economice.
Answer (8 May 2025, 11:50):
Buna ziua. Propunerea nu poate fi aprobată, întrucât se impune conformitatea integrală cu parametrii prevăzuți inițial. O zi buna !
31. Compresor aer cu 6 pistoane
Date:
7 May 2025, 17:00
Question's name:
Solicitare de excludere a unei specificații tehnice
Question:
În cadrul documentației de atribuire, la secțiunea "Specificațiile Tehnice", se menționează următoarea cerință: „Confirmare de la producătorul Diplomat precum că modelul de compresor ofertat este acceptat în lucru cu uniturile existente în instituție, inginer specializat în deservirea utilajului dat.”. Dorim sa solicitam excluderea acestei cerințe, întrucat ea restrange nejustificat concurența și limiteaza participarea altor operatori economici care pot furniza produse echivalente din punct de vedere calitativ și funcțional.Specificația tehnica menționata contravine principiilor transparenței privind achizițiile publice.
Answer (8 May 2025, 11:25):
Buna ziua. Va rugam mult sa va conformati cerintelor Autoritatii Contractante O zi buna !
Date:
12 May 2025, 09:58
Question's name:
Lotul nr. 73 - 73. Aparat portabil de ultrasonografie
Question:
In licitatia precedenta care era similara celei anuntate repetat, au fost date o serie de clarificari la care dvs ati raspuns ca vor for operate modificari in caietul de sarcini. Cu toate acestea, dupa anularea licitatei precedente si anuntarea licitatiei repetate, Solicitam repetat sa adimiteti o erroare de 0.5- 1 MHz, la cetintele tehnice din caietul de sarcini, care ati zis ca va fi admisa in clarificarile din licitatia trecuta
Answer (12 May 2025, 14:37):
Corectat. Mulțumim pentru colaborare.
75. Analizator semiautomat de biochimie
Date:
12 May 2025, 11:24
Question's name:
Descrierea
Question:
Au fost prezentate toate argumentele și au fost furnizate toate referințele, care arată că descrierea aparatului solicitat a fost preluată dintr-un singur model, ceea ce nu permite propunerea altor aparate din partea agenților economici, deoarece propunerea va fi respinsă din cauza nerespectării cerințelor tehnice. Solicităm cu fermitate revizuirea întregii descrieri a dispozitivului solicitat și anularea acestui lot, deoarece descrierea furnizată de dvs. se regăsește doar la un singur producător, acordându-i astfel prioritate, ceea ce este complet contrar principiilor licitației. Chiar și în condițiile în care nu specificați denumirile comerciale și producătorii, în urma analizei ajungem la concluzia că o singură variantă este potrivită, ceea ce arată indirect că descrierea generală nu este respectată. Consultați-vă din nou cu laboratorul și juriștii, analizați descrierea generală de la CAPSC. În caz contrar, acest motiv va duce la o sesizare la ANSC. Observați că acest fapt este observat și la alte loturi din această licitație. Vă mulțumim mult pentru înțelegere Dvs.
Answer (12 May 2025, 11:45):
Buna ziua. Specificațiile tehnice sunt conform necesităților Autorității contractante.
Date:
13 May 2025, 18:27
Question's name:
Lot 31. Compresor aer cu 6 pistoane
Question:
Propunem revizuirea cerintelor tehnice solicitate, deoarece fraza (-Confirmare de la producătorul Diplomat precum că modelul de compresor ofertat este acceptat în lucru cu uniturile existente în instituție), , este una limitativa si neargumentata. Pentru buna functionare a scaunelor este important sa fie asigurat fluxul de aer necesar si presiunea .
Date:
13 May 2025, 18:34
Question's name:
Caiet de sarcini Lot 31
Question:
Cerintele pentru acest lot sunt incomplete si necesita revizuire. Nu s-a indicat de care tip trebuie sa fie compresorul - Cu ulei sau fara ulei. S-au solicitat 3 motoare cate 2 pistoane in timp ce pe piata sunt compresoare de tip cu surub sau cu piston care cu un singur motor pot oferi o productivitate mai mare. Nu s-au luat in considerare sistema de uscare a aerului, valva de purjare automata a condensatului care se acumuleaza in butelie si filtrele de aer care opresc toate impuritatile . Toate aceste sisteme sunt indispensabile pentru buna functionare a scaunelor stomatologice si a tuturor pieselor pneumatice.
Question's name