Status Purchase has not taken place
Estimated value without VAT 700 000 MDL
Period of clarifications: 24 Apr 2024, 17:20 - 17 May 2024, 8:02
Submission of proposals: 17 May 2024, 8:02 - 6 Jun 2024, 8:02
Auction start date: 7 Jun 2024, 9:18

Supplier technical support:

(+373) 79999801

Lot nr. 2 Monitor de pacient compatibil cu IRM 3Tesla
Information about customer
Fiscal code/IDNO
Address
MD-2009, MOLDOVA, mun.Chişinău, mun.Chişinău, MD-2009, Republica Moldova, Chişinău str. Cosmescu 3
Web site
---
The contact person
Full name
Sergiu Balta
Contact phone
+ 373 022 222 364
Purchase data
Date created
24 Apr 2024, 17:16
Date modified
10 May 2024, 13:36
Estimated value (without VAT)
700 000 MDL
The minimum downward of the price
7 MDL
Achizitii.md ID
21220841
Type of procedure
Open tender
Award criteria
The lowest price
Delivery address
MD-2009, MOLDOVA, mun.Chişinău, mun.Chişinău, MD-2009, Republica Moldova, Chişinău str. Cosmescu 3
Contract period
3 Jun 2024 03:00 - 31 Dec 2024 02:00
List of positions
1)
Title
Lot nr. 2 Monitor de pacient compatibil cu IRM 3Tesla CPV: 33100000-1 - Medical equipments
Quantity: 1.0
Unit of measurement: Bucata
Documents of the procurement procedure
contract nou.docx
Bidding Documents
-
24.04.24 17:20
ds_bunuri_servicii_omf_115_15_09_2021.docx
Bidding Documents
-
24.04.24 17:20
duae_ro_0 (1).doc
Bidding Documents
-
24.04.24 17:20
formularele_
Bidding Documents
-
10.05.24 13:36
The history of the auction
View auction
Lot nr. 2 Monitor de pacient compatibil cu IRM 3Tesla
Date:
3 May 2024, 17:50
Question's name:
Monitor pacient
Question:
Unii producători de monitoare de pacient limitează distanța de utilizare, de exemplu - nu mai aproape de 0,5 metri de magnet. Sistemele RMN 3 Tesla au o acțiune foarte puternică asupra obiectelor metalice / electronice. Pentru a nu prezenta pericol la utilizarea acestor sisteme, este mai sigur includerea cerinței să nu existe limită de distanță în utilizarea monitorului de pacient, acesta fiind complet neferomagnetic.
Answer (10 May 2024, 10:49):
Distanța de utilizare nu mai aproape de 0,5 metri este acceptabilă pentru exploatarea monitorului în zona mesei pacientului.
Unfortunately, questions can be asked only during the "Active" period.